Application
UCL-APA301/401 serial burst/EFT pulse verification kit is usually used to the periodic verification of the test system such as
burst generator and immunity test analyzer etc., In order to ensure good functionality and accuracy. MS301/401 consists of
two terminations with 50 Ohms and 1000 Ohms load impedance each. These terminations are rated for up to 8.0KV
burst/EFT pulses and designed to comply with GB/T_17626.4-2008 and the new requirements of IEC/EN 61000-4-4:
amendment 2: 2001. The verification kit can be used To verify almost all brands of burst/FET Generator with kinds of
adaptors such as SHV, BNC, SOV, DN etc.
Specification
Specifications |
UCL-APA 301-4000 |
UCL-APA 301-8000 |
UCL-APA 401-4000 |
UCL-APA 401-8000 |
Max power rating |
3W rms |
4W rms |
Max peak voltage |
4kV |
8kV |
4kV |
8kV |
Max operating time |
120 s (5kHz,300ms/15ms) |
300 s (5kHz,300ms/15ms) |
Nominal attenuation |
-54dB or -60dB |
-60dB |
Nominal voltage radio |
500:1(equal to -54dB)
1000:1(equal to -60dB) |
1000:1 |
Input impedance |
50Ω |
1000Ω |
Output impedance |
50Ω |
50Ω |
Bandwidth (3dB) |
400MHz |
400MHz |
Connectors Input |
SHV,BNC,SOV,DN |
SHV,BNC,SOV,DN |
Output |
BNC female |
BNC female |
Temperature Range |
-10°C~+40°C |
Dimensions |
131.6×25.5×25.5mm |
Weight |
140g |